Optimization of acquisition orbits in cone-beam brain SPECT

Hongwei Ye1, Levon Vogelsang1, David Feiglin2 and Andrzej Krol2

1 Radiology, SUNY Upstate Medical University, Syracuse, New York; 2 Physics, Syracuse University, Syracuse, New York

1678

Objectives: To improve contrast-to-noise ratio for brain imaging with constrain imposed by the total scan duration, and elimination of sampling and truncation artifacts using custom-made cone-beam collimators (CBCs).

Methods: Four different acquisition orbits have been studied for dual-headed gamma camera with two CBCs: single-circle, helical, helix-plus-arc, and CBC combined with parallel-beam collimator (PCB) in a single circular orbit. Analytical projection datasets were created for a numerical mini-Defrise phantom in a 64x64 matrix with CBC focal length = 70 cm and the radius of rotation = 32 cm. Furthermore, Tc-99m HMPAO brain SPECT scan was simulated by Monte Carlo package (SimSET) for the numerical Zubal brain phantom with the same collimators. The detector response, scatter and Poisson noise were considered in Monte Carlo simulations, but not in analytical projections. All projections were reconstructed using ordered subsets expectation maximization with a volumetric system model and exact attenuation correction (OS = 5, up to 100 iterations).

Results: Both qualitative and quantitative evaluations indicate that axial distortion artifacts are always found in reconstructed images acquired in the single-circle orbit since it cannot satisfy the Tuy’s sampling sufficiency condition. These artifacts could be suppressed by using other orbits. However, the helix-plus-arc orbit requires a large number of views ergo long imaging time and more iterations to eliminate artifacts.

Conclusions: For an object with the relatively short axial extent (such as human brain) and CBC with medium focal length (>50 cm), the improved contrast-to-noise and artifacts-free reconstructed images are obtained in SPECT with helical orbit with two CBCs or in combined CBC-PBC SPECT with a single circle orbit.





This Article
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Google Scholar
Right arrow Articles by Ye, H.
Right arrow Articles by Krol, A.
PubMed
Right arrow Articles by Ye, H.
Right arrow Articles by Krol, A.